A Magnetic Electron Analyzer for Plasma Focus Electron Energy Distribution Studies
The paper describes the construction of a magnetic electron analyzer for pulsed beam electron energy distribution studies. The single shot information is obtained using a NMOS linear image sensor. Both the energetical and the sensitivity calibration were performed using careful numerical simulations...
Gespeichert in:
Veröffentlicht in: | Journal of fusion energy 2006-06, Vol.25 (1-2), p.57-66 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The paper describes the construction of a magnetic electron analyzer for pulsed beam electron energy distribution studies. The single shot information is obtained using a NMOS linear image sensor. Both the energetical and the sensitivity calibration were performed using careful numerical simulations. For simplicity, the signals can be read on an oscilloscope, and they are transmitted via an optic fiber, which allows the analyzer to work while connected to high voltage. The analyzer was successfully implemented on a 15 kV/3 kJ plasma focus device and was used to study the electron energy distribution in the 30–660 keV range. |
---|---|
ISSN: | 0164-0313 1572-9591 |
DOI: | 10.1007/s10894-006-9005-x |