Proof of concept testing of a positive reference material for in vivo and in vitro sensitization testing of medical devices

In vivo skin sensitization tests are required to evaluate the biological safety of medical devices in contact with living organisms to provide safe medical care to patients. Negative and positive reference materials have been developed for biological tests of cytotoxicity, implantation, hemolysis, a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomedical materials research. Part B, Applied biomaterials Applied biomaterials, 2024-02, Vol.112 (2), p.e35386
Hauptverfasser: Okamoto, Yusuke, Fukui, Chie, Kobayashi, Toshio, Morioka, Hisako, Mizumachi, Hideyuki, Inomata, Yoriko, Kaneki, Atsushi, Okada, Masayuki, Haishima, Yuji, Yamamoto, Eiichi, Nomura, Yusuke
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In vivo skin sensitization tests are required to evaluate the biological safety of medical devices in contact with living organisms to provide safe medical care to patients. Negative and positive reference materials have been developed for biological tests of cytotoxicity, implantation, hemolysis, and in vitro skin irritation. However, skin sensitization tests are lacking. In this study, polyurethane sheets containing 1 wt/wt % 2,4-dinitrochlorobenzene (DNCB-PU) were developed and evaluated as a positive reference material for skin sensitization tests. DNCB-PU sheet extracts prepared with sesame oil elicited positive sensitization responses for in vivo sensitization potential in the guinea pig maximization test and the local lymph node assay. Furthermore, DNCB-PU sheet extracts prepared with water and acetonitrile, 10% fetal bovine serum-containing medium, or sesame oil elicited positive sensitization responses as alternatives to animal testing based on the amino acid derivative reactivity assay, human cell line activation test, and epidermal sensitization assay, respectively. These data suggest that the DNCB-PU sheet is an effective extractable positive reference material for in vivo and in vitro skin sensitization testing in medical devices. The formulation of this reference material will lead to the development of safer medical devices that contribute to patient safety.