Accuracy of additively manufactured and steam sterilized surgical guides by means of continuous liquid interface production, stereolithography, digital light processing, and fused filament fabrication

Different printing technologies can be used for prosthetically oriented implant placement, however the influence of different printing orientations and steam sterilization remains unclear. In particular, no data is available for the novel technology Continuous Liquid Interface Production. The object...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the mechanical behavior of biomedical materials 2024-04, Vol.152, p.106418-106418, Article 106418
Hauptverfasser: Burkhardt, Felix, Handermann, Leon, Rothlauf, Severin, Gintaute, Aiste, Vach, Kirstin, Spies, Benedikt C., Lüchtenborg, Jörg
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Different printing technologies can be used for prosthetically oriented implant placement, however the influence of different printing orientations and steam sterilization remains unclear. In particular, no data is available for the novel technology Continuous Liquid Interface Production. The objective was to evaluate the dimensional accuracy of surgical guides manufactured with different printing techniques in vertical and horizontal printing orientation before and after steam sterilization. A total of 80 surgical guides were manufactured by means of continuous liquid interface production (CLIP; material: Keyguide, Keyprint), digital light processing (DLP; material: Luxaprint Ortho, DMG), stereolithography (SLA; Surgical guide, Formlabs), and fused filament fabrication (FFF; material: Clear Base Support, Arfona) in vertical and horizontal printing orientation (n = 10 per subgroup). Spheres were included in the design to determine the coordinates of 17 reference points. Each specimen was digitized with a laboratory scanner after additive manufacturing (AM) and after steam sterilization (134 °C). To determine the accuracy, root mean square values (RMS) were calculated and coordinates of the reference points were recorded. Based on the measured coordinates, deviations of the reference points and relevant distances were calculated. Paired t-tests and one-way ANOVA were applied for statistical analysis (significance p  0.05). Regarding reference points and distances, CLIP showed larger deviations compared to SLA in both printing orientations after steam sterilization, while DLP manufactured guides were the most dimensionally stable. In conclusion, the different printing technologies and orientations had little effect on the manufacturing accuracy of the surgical guides before sterilization. However, after sterilization, FFF surgical guides exhibited significant deformation making their clinical use impossible. CLIP showed larger deformations due to steam sterilization than the other photopolymerizing techniques, however, discrepancies may be considered within the range of clinical acceptance. The influence on the implant position rem
ISSN:1751-6161
1878-0180
DOI:10.1016/j.jmbbm.2024.106418