Performance evaluation of Gallium-68 radiopharmaceuticals production using liquid target PETtrace 800 cyclotron

Due to increased demand, cyclotron has an expanding role in producing Gallium-68 (68Ga) radiopharmaceuticals using solid and liquid targets. Though the liquid target produces lower end-of-bombardment activity compared to the solid target, our study presents the performance of 68Ga radiopharmaceutica...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied radiation and isotopes 2024-03, Vol.205, p.111161-111161, Article 111161
Hauptverfasser: Ashhar, Zarif, Ahmad Fadzil, Muhammad Fakhrurazi, Md Safee, Zaitulhusna, Aziz, Firdaus, Ibarhim, Ummi Habibah, Nik Afinde, Nik Muhammad Fitri, Mat Ail, Noratikah, Jamal Harizan, Muhammad Ali Hazizi, Halib, Dzulieza, Alek Amran, Arieffuddin, Adawiyah, Rabiatul, Abd Hamid, Mohd Hamdi Noor, Mahamood, Mazurin, Razali, Nor Idayu, Said, Mohamad Aminudin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Due to increased demand, cyclotron has an expanding role in producing Gallium-68 (68Ga) radiopharmaceuticals using solid and liquid targets. Though the liquid target produces lower end-of-bombardment activity compared to the solid target, our study presents the performance of 68Ga radiopharmaceuticals production using the liquid target by evaluating the end-of-bombardment activity and the end-of-purification activity of [68Ga]GaCl3. We also present the effect of increasing irradiation time, which significantly improves the end-of-synthesis yield. From the result obtained, the end-of-bombardment activity produced was 4.48 GBq, and the [68Ga]GaCl3 end-of-purification activity produced was 2.51 GBq with below-limit metallic impurities. Increasing the irradiation time showed a significant increase in the end-of-synthesis activity from 1.33 GBq to 1.95 GBq for [68Ga]Ga-PSMA-11 and from 1.13 GBq to 1.74 GBq for [68Ga]Ga-DOTA-TATE. Based on the improvements made, the liquid target production of 68Ga radiopharmaceuticals is feasible and reproducible to accommodate up to 5 patients per production. In addition, this work also discusses the issues encountered, together with the possible corrective and preventative measures. [Display omitted] •More than 1.7 GBq [68Ga]Ga-PSMA and [68Ga]Ga-DOTATATE is produced using liquid target.•Increasing the target solution's HNO3 concentration and the irradiation time improves the End of Synthesis (EOS) yield.•Consistent production of 68Ga-radiopharmaceuticals can be achieved through proper maintenance and preventative measures.
ISSN:0969-8043
1872-9800
DOI:10.1016/j.apradiso.2023.111161