Detectability of cold tumors by xSPECT bone technology compared with hot tumors: a supine phantom study
Detecting cold as well as hot tumors is vital for interpreting bone tumors on single-photon emission computed tomography (SPECT) images. This study aimed to visually and quantitatively demonstrate the detectability of cold tumors using xSPECT technology compared with that of hot tumors in the phanto...
Gespeichert in:
Veröffentlicht in: | Australasian physical & engineering sciences in medicine 2024-03, Vol.47 (1), p.287-294 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Detecting cold as well as hot tumors is vital for interpreting bone tumors on single-photon emission computed tomography (SPECT) images. This study aimed to visually and quantitatively demonstrate the detectability of cold tumors using xSPECT technology compared with that of hot tumors in the phantom study. Five tumors of different sizes and normal bone contained a mixture of
99m
Tc and K
2
HPO
4
in a spine phantom. We acquired SPECT data using an xSPECT protocol and transverse images were reconstructed using xSPECT Bone (xB) and xSPECT Quant (xQ). Mean standardized uptake values (SUV
mean
) in volumes of interest (VOI) were calculated. Recovery coefficients (RCs) for each tumor site were calculated with reference to radioactive concentrations. The SUV
means
of the whole vertebral body for hot tumor bone image in cortical bone phantom reconstructed by with xB and xQ were 5.77 and 4.86 respectively. The SUV
mean
of xB was similar to the true value. The SUV
means
for xB and xQ reconstructed images of cold tumors were both approximately 0.16. The RC of the cold tumor on xQ images increased as the tumor diameter decreased, whereas that of xB remained almost constant regardless of the tumor diameter. In conclusion, the quantitative accuracy of detecting hot and cold tumors was higher in the xB image than in the xQ image. Moreover, the visual detectability of cold tumors was also excellent in xB images. |
---|---|
ISSN: | 2662-4729 0158-9938 2662-4737 1879-5447 |
DOI: | 10.1007/s13246-023-01364-y |