Validity and Reliability of Upper Limb Kinematic Assessment Using a Markerless Motion Capture (MMC) System: A Pilot Study
To investigate the validity and test-retest reliability of a customized markerless motion capture (MMC) system that used iPad Pros with a Light Detection And Ranging scanner at two different viewing angles to measure the active range of motion (AROM) and the angular waveform of the upper-limb-joint...
Gespeichert in:
Veröffentlicht in: | Archives of physical medicine and rehabilitation 2024-04, Vol.105 (4), p.673-681.e2 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To investigate the validity and test-retest reliability of a customized markerless motion capture (MMC) system that used iPad Pros with a Light Detection And Ranging scanner at two different viewing angles to measure the active range of motion (AROM) and the angular waveform of the upper-limb-joint angles of healthy adults performing functional tasks.
Participants were asked to perform shoulder and elbow actions for the investigator to take AROM measurements, followed by four tasks that simulated daily functioning. Each participant attended 2 experimental sessions, which were held at least 2 days and at most 14 days apart.
A Vicon system and 2 iPad Pros installed with our MMC system were placed at 2 different angles to the participants and recorded their movements concurrently during each task.
Thirty healthy adults (mean age: 28.9, M/F ratio: 40/60).
Not applicable.
The AROM and the angular waveform of the upper-limb-joint angles.
The iPad Pro MMC system underestimated the shoulder joint and elbow joint angles in all four simulated functional tasks. The MMC demonstrated good to excellent test-retest reliability for the shoulder joint AROM measurements in all 4 tasks.
The maximal AROM measurements calculated by the MMC system had consistently smaller values than those measured by the goniometer. An MMC in iPad Pro system might not be able to replace conventional goniometry for clinical ROM measurements, but it is still suggested for use in home-based and telerehabilitation training for intra-subject measurements because of its good reliability, low cost, and portability. Further development to improve its performance in motion capture and analysis in disease populations is warranted. |
---|---|
ISSN: | 0003-9993 1532-821X 1532-821X |
DOI: | 10.1016/j.apmr.2023.10.018 |