Wear of Third-Generation Cross-Linked Polyethylene in Primary Total Hip Arthroplasty: A 10-Year Analysis
Sequential modifications to the manufacturing process of highly cross-linked polyethylene (HXLPE) have improved the wear resistance and implant survival of these liners in total hip arthroplasty (THA). However, no study has examined the long-term (mean 10 year) wear rates and clinical outcomes of th...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2024-05, Vol.39 (5), p.1335-1340 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Sequential modifications to the manufacturing process of highly cross-linked polyethylene (HXLPE) have improved the wear resistance and implant survival of these liners in total hip arthroplasty (THA). However, no study has examined the long-term (mean 10 year) wear rates and clinical outcomes of third-generation HXLPE in THA. The aim of our study was to report the longest-to-date analysis of wear rates and clinical outcomes of a third-generation HXLPE liner.
A series of 133 THAs using a specific HXLPE acetabular liner were retrospectively evaluated. Linear and volumetric wear rates were determined using a validated radiographic technique and clinical outcomes were analyzed. Multivariate analyses were performed to determine risk factors for accelerated wear.
At a mean follow-up of 10.4 years (range, 8 to 13.4), the mean linear wear rate was 0.0172 mm/year and the mean volumetric wear rate was 16.99 mm3/year. There were no instances of osteolysis or mechanical failures at any time point and there was a 100% acetabular component survival rate. Younger age and use of offset liners were independent risk factors for increased wear (P < .01 for both).
Our series of a third-generation HXLPE demonstrated very low wear rates and excellent implant survival at a mean of 10.4 years following primary THA. Future comparative studies at the 15- and 20-year follow-up timepoints are necessary to determine if such findings translate to true improvements in the tribological properties and longevity of these liners when compared to previous generations of HXLPE liners. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2023.11.016 |