Test-retest repeatability and agreement of the quantitative contrast sensitivity function test: towards the validation of a new clinical endpoint

Purpose The purpose of this study is to investigate test-retest reliability and agreement of the quantitative contrast sensitivity function test (qCSF) in the retina clinic. Methods A total of 121 right eyes of 121 patients were tested and consecutively re-tested with qCSF in the retina clinic. Outc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Graefe's archive for clinical and experimental ophthalmology 2024-03, Vol.262 (3), p.813-822
Hauptverfasser: Finn, Matthew, Vingopoulos, Filippos, Zhao, Yan, Zhou, Paul, Bannerman, Augustine, Romano, Francesco, Ding, Xinyi, Hassan, Zakariyya, Patel, Nimesh A., Wu, David M., Miller, John B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose The purpose of this study is to investigate test-retest reliability and agreement of the quantitative contrast sensitivity function test (qCSF) in the retina clinic. Methods A total of 121 right eyes of 121 patients were tested and consecutively re-tested with qCSF in the retina clinic. Outcomes included area under the logarithm of contrast sensitivity function curve (AULCSF), contrast acuity, and contrast sensitivity thresholds at 1–18 cycles per degree (cpd). Test-retest means were compared with paired t -test, variability was compared with the Brown-Forsythe test, and intraclass correlation coefficient (ICC) and Bland Altman plots evaluated reliability and agreement. Results Mean test-retest differences for all qCSF metrics ranged from 0.02 to 0.05 log units without statistically significant differences in variability. Standard deviations ranged from 0.08 to 0.14. Coefficients of repeatability ranged from 0.16 to 0.27 log units. ICC > 0.9 for all metrics except 1cpd (ICC = 0.84, all p < 0.001); AULCSF ICC = 0.971. Conclusion qCSF-measured contrast sensitivity shows great test-retest repeatability and agreement in the retina clinic.
ISSN:0721-832X
1435-702X
DOI:10.1007/s00417-023-06291-y