Reconstruction of the proximal radius with 3D-printed personalized prosthesis after tumor resection: case series
Giant cell tumor of bone (GCTB) (Campanacci III) or malignant tumors extend to the epiphyseal region of the proximal radius, and intra-articular resection of the proximal radius is often needed. In the present study, we present the patients who underwent reconstruction of the proximal radius with 3D...
Gespeichert in:
Veröffentlicht in: | Journal of shoulder and elbow surgery 2024-03, Vol.33 (3), p.556-563 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Giant cell tumor of bone (GCTB) (Campanacci III) or malignant tumors extend to the epiphyseal region of the proximal radius, and intra-articular resection of the proximal radius is often needed. In the present study, we present the patients who underwent reconstruction of the proximal radius with 3D-printed personalized prosthesis after tumor resection, aiming to describe the prosthesis design and surgical technique and evaluate the clinical outcomes of this method.
Between November 2018 and January 2021, 9 patients received radial hemiarthroplasty with 3D-printed personalized prostheses after tumor resection. The pathologic diagnosis was GCTB (Campanacci III) in 7 patients, osteosarcoma (IIB) in 1 patient, and synovial sarcoma (IIB) in 1 patient. The range of motion (ROM) and strength in terms of elbow flexion/extension and forearm supination/pronation were evaluated. Pain was assessed by the visual analog scale (VAS) preoperatively and at each follow-up visit. To evaluate the functional outcome, the Mayo Elbow Performance Score (MEPS) system and the Musculoskeletal Tumor Society (MSTS) scoring system were administered at each follow-up visit. Complications and oncological outcomes were recorded.
The patients were followed from 24 to 51 months, with a median follow-up of 35 months. No patients were lost to follow-up. During the follow-up, local recurrence and metastasis were not observed. The VAS score improved from a median of 5 points (range 4-7) preoperatively to 1 point (range 0-2) at the last follow-up visit. The mean MEPS score was 88.5% (83-93), and the mean MSTS score was 25.3 (24-27) at the last follow-up visit. No complications such as infection and aseptic loosening were detected.
The implantation of a 3D-printed personalized prosthesis after proximal radial resection showed excellent oncologic outcomes and postoperative function at short-term follow-up and is a viable alternative method for reconstruction of the proximal radius bone defect after tumor resection. |
---|---|
ISSN: | 1058-2746 1532-6500 |
DOI: | 10.1016/j.jse.2023.08.032 |