Trueness and precision of combined healing abutment scan body system scans at different sites of maxilla after multiple repositioning of the scan body
To evaluate the accuracy of the scans of the combined healing abutment-scan body (CHA-SB) system located at different sites of the maxilla when SBs are replaced in between each scan. Three SBs were seated into HAs located at the central incisor, first premolar, and first molar sites of a maxillary m...
Gespeichert in:
Veröffentlicht in: | Journal of dentistry 2023-12, Vol.139, p.104716-104716, Article 104716 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the accuracy of the scans of the combined healing abutment-scan body (CHA-SB) system located at different sites of the maxilla when SBs are replaced in between each scan.
Three SBs were seated into HAs located at the central incisor, first premolar, and first molar sites of a maxillary model inside a phantom head, and the model was scanned extraorally (CEREC Primescan SW 5.2). This procedure was repeated with new SBs until a total of 10 scans were performed. Standard tessellation language files of CHA-SBs at each implant location were isolated, transferred into analysis software (Geomagic Control X), and superimposed over the proprietary library files to analyze surface (root mean square), linear, and angular deviations. Trueness and precision were evaluated with one-way analysis of variance and Tukey tests. The correlation between surface and angular deviations was analyzed with Pearson's correlation (α=0.05).
Molar implant scans had the highest surface and angular deviations (P≤.006), while central incisor implant scans had higher precision (surface deviations) than premolar implant scans (P=.041). Premolar implant scans had higher accuracy than central incisor implant scans on the y-axis (P≤.029). Central incisor implant scans had the highest accuracy on the z-axis (P≤.018). A strong positive correlation was observed between surface and angular deviations (r = 0.864, P |
---|---|
ISSN: | 0300-5712 1879-176X |
DOI: | 10.1016/j.jdent.2023.104716 |