Is the automated Elecsys tacrolimus assay on the Roche cobas e 602 analyzer an acceptable replacement for a liquid chromatography–tandem mass spectrometry–based assay?
Abstract Objectives This study evaluated the suitability and accuracy of the automated Roche Elecsys tacrolimus electrochemiluminescence immunoassay (ECLIA) by comparing it with a current laboratory-developed test by liquid chromatography–tandem mass spectrometry (LC-MS/MS). Methods The tacrolimus E...
Gespeichert in:
Veröffentlicht in: | American journal of clinical pathology 2024-01, Vol.161 (1), p.97-106 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Objectives
This study evaluated the suitability and accuracy of the automated Roche Elecsys tacrolimus electrochemiluminescence immunoassay (ECLIA) by comparing it with a current laboratory-developed test by liquid chromatography–tandem mass spectrometry (LC-MS/MS).
Methods
The tacrolimus ECLIA was evaluated for precision, linearity, interference, and postextraction stability. Accuracy was compared with LC-MS/MS.
Results
The tacrolimus ECLIA assay is precise, exhibits a measuring range of 0.75 to 30 ng/mL, and is tolerant of significant interferences (plasma indices: hemolysis |
---|---|