A new way to visualize prostate brachytherapy needles using ultrasound color Doppler and needle surface modifications

Suboptimal ultrasound conspicuity of the brachytherapy applicator can lead to inaccurate image reconstructions of the applicator resulting in decreased tumor control or increased normal tissue dose. This feasibility study aims to improve ultrasound conspicuity of high-dose rate (HDR) brachytherapy n...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brachytherapy 2023-11, Vol.22 (6), p.761-768
Hauptverfasser: Dupere, Justine M., Brost, Eric E., Uthamaraj, Susheil, Lee, Christine U., Urban, Matthew W., Deufel, Christopher L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Suboptimal ultrasound conspicuity of the brachytherapy applicator can lead to inaccurate image reconstructions of the applicator resulting in decreased tumor control or increased normal tissue dose. This feasibility study aims to improve ultrasound conspicuity of high-dose rate (HDR) brachytherapy needles by modifying the surface of the needles to produce a color Doppler twinkling signature. Surface modifications of standard 17-gauge titanium HDR brachytherapy needles included laser-scribing, application of polymethyl methacrylate (PMMA), and coating with a commercially available echogenic coating. Laser-scribing was performed with variable widths (0.1-1 mm) and depths (10-100 µm). The echogenic coating was applied with three different thicknesses (27, 40, and 64 µm). Unmodified and modified needles were imaged under B-mode and color Doppler ultrasound in phantom and cadaver, and the signal strength was recorded. Laser-scribed, PMMA-coated, and echogenic-coated brachytherapy needles produced a twinkling signature along the needle shaft on color Doppler ultrasound. Twinkling was observed with laser-scribe depths >20 µm and widths >0.1 mm and from echogenic coatings 40 µm and 64 µm thick. Twinkling was not observed with unmodified needles. The twinkling signature had a spectral composition with a uniform magnitude between the velocities of 2 to 16 cm/s. Color Doppler ultrasound of surface-modified brachytherapy applicators may improve applicator conspicuity aiding applicator placement and digitization. HDR brachytherapy needles may be modified to produce the twinkling signature via laser-scribing, PMMA rings, or applying an echogenic coating.
ISSN:1538-4721
1873-1449
DOI:10.1016/j.brachy.2023.07.002