Tribological behaviour of plasma nitrided Ti-5Al-2Nb-1Ta alloy against UHMWPE

A detailed study has been made on the wear behaviour of untreated and plasma nitrided Ti-5Al-2Nb-1Ta orthopaedic alloy against ultra high molecular weight polyethylene (UHMWPE) using pin on disc tribometer under lubricated conditions. The effects of nitriding temperature and nitriding time on the ba...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Tribology international 2004-08, Vol.37 (8), p.627-631
Hauptverfasser: Gokul Lakshmi, S, Arivuoli, D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A detailed study has been made on the wear behaviour of untreated and plasma nitrided Ti-5Al-2Nb-1Ta orthopaedic alloy against ultra high molecular weight polyethylene (UHMWPE) using pin on disc tribometer under lubricated conditions. The effects of nitriding temperature and nitriding time on the basis of the evolution of the wear volume loss and friction coefficient were investigated. The wear resistance of the plasma nitrided alloys increased considerably when compared to the untreated alloy. The wear debris identified using X-ray diffraction measurements indicated the formation of titanium oxide and titanium oxynitride particles. The wear rate was found to increase with increase in load and sliding velocity.
ISSN:0301-679X
1879-2464