Intercomparison of NIST, NPL, PTB, and VSL thermal voltage converters from 100 kHz to 1 MHz

Coaxial, thermal voltage converters were hand-carried among NIST, NPL, PTB, and VSL for comparison of AC-DC difference from 100 kHz to 1 MHz. The highly varied methods and underlying principles on which AC-DC difference determinations are based in each laboratory, the transport standards used, and t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 1993-04, Vol.42 (2), p.615-617
Hauptverfasser: Kinard, J.R., Knight, R.B.D., Martin, P., Klonz, M., de Vreede, J.P.M., Dessens, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Coaxial, thermal voltage converters were hand-carried among NIST, NPL, PTB, and VSL for comparison of AC-DC difference from 100 kHz to 1 MHz. The highly varied methods and underlying principles on which AC-DC difference determinations are based in each laboratory, the transport standards used, and the results of the comparisons are presented. The AC-DC differences reported by the participating laboratories are in very good agreement. It is concluded that representative values of AC-DC difference in the 100 kHz to 1 MHz frequency range are established with 2 sigma uncertainties of 4-8 ppm.< >
ISSN:0018-9456
1557-9662