Uncertainty analysis of the transmission-type measurement of Q-factor
The transmission-type measurement of the Q-factor of a microwave resonator is analyzed with a goal of establishing uncertainty limits on the unloaded Q-factor. The three major sources of systematic errors which are discussed are: the instrumentation inaccuracy, the inequality of the input and output...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on microwave theory and techniques 1999-03, Vol.47 (3), p.367-371 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The transmission-type measurement of the Q-factor of a microwave resonator is analyzed with a goal of establishing uncertainty limits on the unloaded Q-factor. The three major sources of systematic errors which are discussed are: the instrumentation inaccuracy, the inequality of the input and output coupling coefficient, and the effect of coupling losses. It is shown that, for tightly coupled resonators, the uncertainty of the transmission-type method is significantly increased. |
---|---|
ISSN: | 0018-9480 1557-9670 |
DOI: | 10.1109/22.750244 |