The corrosion–wear behaviour of thermally oxidised CP-Ti and Ti–6Al–4V
The use of commercial purity titanium (CP-Ti) and Ti–6Al–4V alloys in bio-medical implant applications has been limited by their poor resistance to surface degradation processes. In this paper the corrosion–wear behaviour of untreated and thermally oxidised CP-Ti and Ti–6Al–4V have been compared. Ox...
Gespeichert in:
Veröffentlicht in: | Wear 2004-03, Vol.256 (5), p.469-479 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The use of commercial purity titanium (CP-Ti) and Ti–6Al–4V alloys in bio-medical implant applications has been limited by their poor resistance to surface degradation processes. In this paper the corrosion–wear behaviour of untreated and thermally oxidised CP-Ti and Ti–6Al–4V have been compared. Oxidation of both alloys at 625
°C for 36
h resulted in the formation of an exterior layer of TiO
2 (rutile) that had a hardness ∼1000
HV. Corrosion–wear tests were made in reciprocation sliding contact with an α-Al
2O
3 ball immersed in physiological saline (0.89% NaCl) at room temperature. The oxidation treatment retarded the corrosion–wear of both CP-Ti and Ti–6Al–4V. For the untreated alloys, surface damage was dominated by micro-asperity shearing which resulted in rapid wear. Corrosion–wear of the oxidised materials was slower but more complex. The exterior TiO
2 layer formed on the oxidised Ti–6Al–4V alloy provided little protection, it was rapidly removed during the first 60
min of testing, by a process involving interfacial fracture. Conversely, the TiO
2 layer, albeit thinner, provided protection for the oxidised CP-Ti. Here, the layer becomes smoothly worn by a process that is proposed to be caused by the mechanical dissociation of the TiO
2-layer. For both oxidised titanium alloys the hardened oxygen diffusion zone (ODZ), formed beneath the TiO
2 layer, provided good protection from corrosion–wear. In both cases the ODZ was smoothly worn by a combination of abrasion and corrosion–wear processes. The latter process, termed Type I corrosion–wear, involves the repetitive mechanical degradation of the passive film that forms through aqueous corrosion. However, this is a relatively slow process. |
---|---|
ISSN: | 0043-1648 1873-2577 |
DOI: | 10.1016/S0043-1648(03)00557-X |