Generalized adaptive calibration schemes for precision RF vector network analyzer measurements
This paper discusses RF calibration schemes for one-port vector network analyzers other than the traditional short-open-load (SOL) scheme. Before use, the impedance standards are characterized by fitting polynomials to measured voltage reflection coefficient values. Simulated uncertainty profiles an...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 2003-08, Vol.52 (4), p.1266-1272, Article 1266 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper discusses RF calibration schemes for one-port vector network analyzers other than the traditional short-open-load (SOL) scheme. Before use, the impedance standards are characterized by fitting polynomials to measured voltage reflection coefficient values. Simulated uncertainty profiles and example uncertainties of measurement are presented for various calibration schemes. Examples of non-SOL schemes and a method of adaptively choosing the best calibration scheme for a particular device under test at a particular frequency are also presented. |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/TIM.2003.816811 |