An M-503 magnetic-field strength gage
The main concepts of a method for calculating gages for magnetic-field strengths no higher than 500 kA/m are considered. This method is intended for calibrating nondestructive-testing instruments. The M-503 gage can be both ac- and dc-operated. The gage error is within the range from 0.5 to 1.5%. [M...
Gespeichert in:
Veröffentlicht in: | Russian journal of nondestructive testing 2004-11, Vol.40 (11), p.776-781 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The main concepts of a method for calculating gages for magnetic-field strengths no higher than 500 kA/m are considered. This method is intended for calibrating nondestructive-testing instruments. The M-503 gage can be both ac- and dc-operated. The gage error is within the range from 0.5 to 1.5%. [Material tested: electrical steel]. |
---|---|
ISSN: | 1061-8309 1608-3385 |
DOI: | 10.1007/s11181-005-0057-1 |