Accuracy evaluation of the optically pumped Cs frequency standard at NRLM
The optically pumped cesium (Cs) frequency standard at the National Research Laboratory of Metrology, Ibaraki, Japan (NRLM), NRLM-4, was modified in various ways in order to improve frequency stability and accuracy, We determined the frequency stability by comparing the phase with that of a hydrogen...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 1999-04, Vol.48 (2), p.496-499 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The optically pumped cesium (Cs) frequency standard at the National Research Laboratory of Metrology, Ibaraki, Japan (NRLM), NRLM-4, was modified in various ways in order to improve frequency stability and accuracy, We determined the frequency stability by comparing the phase with that of a hydrogen maser: the Allan deviation was /spl sigma//sub y/(/spl tau/)=5/spl times/10/sup -13///spl radic//spl tau/. We estimated the frequency corrections and uncertainties due to various sources, and the total uncertainty was found to be 2.9/spl times/10/sup -14/. The frequency difference between NRLM-4 and the international atomic time (TAI) was also measured. |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/19.769640 |