Use of the Inspiris valve in the native right ventricular outflow tract is associated with early prosthetic regurgitation
The Inspiris Resilia prosthesis (Edwards Lifesciences) has been increasingly used in the pulmonic position with limited performance data. We sought to investigate its durability as a surgical pulmonary valve replacement (PVR). We retrospectively reviewed patients who underwent PVR or conduit replace...
Gespeichert in:
Veröffentlicht in: | The Journal of thoracic and cardiovascular surgery 2023-10, Vol.166 (4), p.1210-1221.e8 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Inspiris Resilia prosthesis (Edwards Lifesciences) has been increasingly used in the pulmonic position with limited performance data. We sought to investigate its durability as a surgical pulmonary valve replacement (PVR).
We retrospectively reviewed patients who underwent PVR or conduit replacement with an Inspiris or non-Inspiris valve/conduit from 2018 to 2022. The primary end point was freedom from a composite of at least moderate pulmonary regurgitation, pulmonary stenosis, or valve/conduit reintervention. Secondary end points were individual components of the composite outcome. To account for baseline differences, propensity matching identified 70 patient pairs.
A total of 227 patients (median age: 19.3 years [interquartile range, 11.8-34.4]) underwent PVR or conduit replacement (Inspiris: n = 120 [52.9%], non-Inspiris: n = 107 [47.1%]). Median follow-up was 26.6 months [interquartile range, 12.4-41.1]. Among matched patients, 2-year freedom from valve failure was lower in the Inspiris group (53.5 ± 9.3% vs 78.5 ± 5.9%, P = .03), as was freedom from at least moderate pulmonary regurgitation (54.2 ± 9.6% vs 86.4 ± 4.9%, P |
---|---|
ISSN: | 0022-5223 1097-685X 1097-685X |
DOI: | 10.1016/j.jtcvs.2023.04.018 |