Prospective multicenter study of a novel endovascular venous anastomotic procedure and device for implantation of an arteriovenous graft for hemodialysis
Introduction: The traditional sutured venous anastomosis used during arteriovenous graft implantation is associated with a high incidence of subsequent stenosis that is attributed to neointimal hyperplasia. Hyperplasia results from multiple factors, including hemodynamic abnormalities and vessel tra...
Gespeichert in:
Veröffentlicht in: | The journal of vascular access 2024-07, Vol.25 (4), p.1244-1251 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction:
The traditional sutured venous anastomosis used during arteriovenous graft implantation is associated with a high incidence of subsequent stenosis that is attributed to neointimal hyperplasia. Hyperplasia results from multiple factors, including hemodynamic abnormalities and vessel trauma during implantation. A novel anastomotic connector device was designed to provide an alternative, less traumatic, endovascular venous anastomosis that may ameliorate the clinical challenges associated with a sutured anastomosis. A prospective single-arm multicenter study was performed to evaluate safety and effectiveness of graft implantation using the study device.
Methods:
Patients requiring graft creation and who met the study criteria were enrolled between February 2018 and July 2021 and observed for 6 months. Collected data included baseline characteristics, graft patency and use for hemodialysis, graft interventions, and adverse events. The primary study endpoint was cumulative graft patency, compared to a pre-specified Performance Goal of 75%. Secondary endpoints included primary unassisted patency and serious adverse events, defined as the occurrence of death, graft infection, emergent surgery, significant bleeding, and pseudoaneurysm.
Results:
A total of 158 patients were enrolled from 10 study sites, among which 144 subjects were evaluable at 6 months and 14 were censored with partial follow-up observation. Three patients died and the graft was abandoned in 12. The primary endpoint was met (p-value |
---|---|
ISSN: | 1129-7298 1724-6032 1724-6032 |