Clinical performance comparison of two medium cut‐off dialyzers

Introduction Medium‐cut‐off (MCO) dialyzers may beneficially impact outcomes in patients on hemodialysis. Methods In a randomized, controlled trial in maintenance hemodialysis patients, the new Nipro ELISIO‐17HX MCO dialyzer was compared to the Baxter Theranova 400 filter regarding middle molecule r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Therapeutic apheresis and dialysis 2023-04, Vol.27 (2), p.284-292
Hauptverfasser: Krieter, Detlef H., Rüth, Marieke, Lemke, Horst‐Dieter, Wanner, Christoph
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction Medium‐cut‐off (MCO) dialyzers may beneficially impact outcomes in patients on hemodialysis. Methods In a randomized, controlled trial in maintenance hemodialysis patients, the new Nipro ELISIO‐17HX MCO dialyzer was compared to the Baxter Theranova 400 filter regarding middle molecule removal. Furthermore, the suitability of two assays for free lambda‐light chain (λFLC) detection (Freelite vs. N‐Latex) was verified. Results ELISIO‐HX achieved slightly lower reduction ratios for β2‐microglobulin (71.8 ± 6.0 vs. 75.3 ± 5.8%; p = 0.001), myoglobin (54.7 ± 8.6 vs. 64.9 ± 8.7%; p 