Chromium measurement in pharmaceutical samples: a comparative study of three new membrane electrodes with different electron-ion exchangers
The present study deals with synthesis of N -(thiazol-2-ylcarbamothioyl) benzamide. It was utilized as a neutral ionophore for designing three types of chromium( iii ) sensors including coated wire ion selective electrodes (CW-ISEs), ion selective electrodes with liquid internal electrolyte (LIE-ISE...
Gespeichert in:
Veröffentlicht in: | Analytical methods 2023-03, Vol.15 (11), p.1431-144 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The present study deals with synthesis of
N
-(thiazol-2-ylcarbamothioyl) benzamide. It was utilized as a neutral ionophore for designing three types of chromium(
iii
) sensors including coated wire ion selective electrodes (CW-ISEs), ion selective electrodes with liquid internal electrolyte (LIE-ISEs), and solid-state ion selective electrodes (SS-ISEs). UV-visible spectrophotometry was used to confirm the affinity of
N
-(thiazol-2-ylcarbamothioyl) benzamide to chromium(
iii
). It was found that a membrane with a composition of 2% NaTPB, 8% ionophore, 60% DBP, and 30% PVC showed the best performance and a Nernstian slope of 21.6 mV per decade. Scanning electron microscopy was used to assess the PVC membrane morphology. The existence of chromium(
iii
) in the liquid membrane matrix was proved by energy-dispersive X-ray spectroscopy. Detection limits for SS-ISE (1 × 10
−6
M) and CW-ISE (1 × 10
−6
M) were enhanced relative to LIE-ISE (1 × 10
−5
M). All three electrodes showed a response time of about 5 s. The sensors' applicable pH range was 4.0-6.0. Fourier transform infrared spectra recorded through the electrode membrane showed that chromium(
iii
) ion can interact with sulfur, nitrogen and oxygen atoms of
N
-(thiazol-2-ylcarbamothioyl) benzamide. The sensors were utilized as indicator electrodes in chromium(
iii
) potentiometric titration with ethylenediaminetetraacetic acid and for directly measuring chromium(
iii
) in some pharmaceutical samples.
Three types of chromium(
iii
) sensors with different electron-ion exchangers were fabricated using the newly synthesized ionophore. The sensors were successfully applied for potentiometric measurement of chromium(
iii
) in pharmaceutical samples. |
---|---|
ISSN: | 1759-9660 1759-9679 |
DOI: | 10.1039/d2ay01601j |