Validation of a blood gas device for ionized calcium analysis in Holstein cows
Background Accurate analysis of ionized calcium (iCa) is critical for the detection of hypocalcemia or subclinical hypocalcemia. The Edan i15 Vet (EDAN) blood gas device has not been validated for iCa in dairy cows. Objectives We aimed to validate the EDAN blood gas device against the Gem Premier 30...
Gespeichert in:
Veröffentlicht in: | Veterinary clinical pathology 2023-06, Vol.52 (2), p.276-283 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Accurate analysis of ionized calcium (iCa) is critical for the detection of hypocalcemia or subclinical hypocalcemia. The Edan i15 Vet (EDAN) blood gas device has not been validated for iCa in dairy cows.
Objectives
We aimed to validate the EDAN blood gas device against the Gem Premier 3000 (GEM) analyzer by measuring iCa concentrations and evaluating the ability of these measurements, compared with serum total calcium (TCa) concentrations, to diagnose subclinical hypocalcemia.
Methods
iCa concentrations were measured with the EDAN and GEM devices, and serum TCa concentrations were measured with a wet biochemistry method with blood from 125 lactating Holstein cows between calving to day 27 postpartum.
Results
Bland‐Altman plots showed a mean and total bias of 0.05 and 0.24 mmol/L for the EDAN device, respectively. The intercept did not include zero, but the slope included 1.0 in the Passing‐Bablok regression. The sensitivity and specificity (Se/Sp) of the EDAN device were 93/94%, 93/90%, 91/93%, and 85/95% for iCa cut‐off values of |
---|---|
ISSN: | 0275-6382 1939-165X |
DOI: | 10.1111/vcp.13233 |