Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model

The purpose of this work is to provide measured data for the modified TG43 parameters [DeWerd et al.] for the newest, Galden-cooled S7600 Xoft Axxent source model. The measurement of radial dose distributions at distances of 1 cm to 4 cm from the source was performed using TLD100 microcubes, EBT3 fi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brachytherapy 2023-03, Vol.22 (2), p.260-268
Hauptverfasser: Walter, Autumn E., Khan, Ahtesham U., DeWerd, Larry A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this work is to provide measured data for the modified TG43 parameters [DeWerd et al.] for the newest, Galden-cooled S7600 Xoft Axxent source model. The measurement of radial dose distributions at distances of 1 cm to 4 cm from the source was performed using TLD100 microcubes, EBT3 film, and an Exradin A26 microionization chamber. The overall uncertainty and reproducibility of each dosimeter was evaluated for its use in determining the radial dose function and dose rate conversion coefficient. An acrylic phantom developed in house for previous works was used to measure the polar anisotropy function using TLD100 microcubes at distances of 1 cm, 2 cm, and 5 cm from the source. The Exradin A26 chamber was deemed most suitable for measuring the radial dose function. Values determined had a maximum k = 1 uncertainty of 1.4%. The dose rate conversion coefficient measured with the chamber was found to be 9.33 ± 0.21cGy/hrμGy/min. TLD100 microcube measurements of the polar anisotropy had average uncertainties of 6%, 3%, and 2.5% at 1 cm, 2 cm, and 5 cm, respectively. The modified TG43 parameters for the bare source were measured with reasonable uncertainty. The values determined will aid with the clinical implementation of the source for breast and endometrial cancer applications.
ISSN:1538-4721
1873-1449
DOI:10.1016/j.brachy.2022.11.010