Amperometric Sensing of Ethylene Oxide in the Gas Phase

The sensing device used is based on a porous Pt electrode, which is supported on an ion-exchange membrane and directly exposed to the gas phase. Under acid conditions, ethylene oxide was found to be oxidized on the platinum oxide surface at +550 mV vs MSE, thus enabling its monitoring via the measur...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2000-05, Vol.72 (10), p.2206-2210
Hauptverfasser: Hodgson, Alexia W. E, Jacquinot, Patrick, Hauser, Peter C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The sensing device used is based on a porous Pt electrode, which is supported on an ion-exchange membrane and directly exposed to the gas phase. Under acid conditions, ethylene oxide was found to be oxidized on the platinum oxide surface at +550 mV vs MSE, thus enabling its monitoring via the measurement of the associated current. A detection limit of 15 ppb was obtained, based on a signal-to-noise ratio of three, and a linear dynamic range was found up to 100 ppm. The effects of mass transport, humidity, and oxygen on the cell response, as well as the cross sensitivity to other organic vapors and inorganic gases are discussed.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac991247l