Initial experience with stylet-driven versus lumenless lead delivery systems for left bundle branch area pacing
Left bundle branch area pacing (LBBP) has emerged as an alternative method for conduction system pacing. While initial experience with delivery systems for stylet-driven and lumenless lead implantation for LBBP has been described, data comparing outcomes of stylet-driven versus lumenless lead implan...
Gespeichert in:
Veröffentlicht in: | Journal of cardiovascular electrophysiology 2023-03, Vol.34 (3), p.710-717 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Left bundle branch area pacing (LBBP) has emerged as an alternative method for conduction system pacing. While initial experience with delivery systems for stylet-driven and lumenless lead implantation for LBBP has been described, data comparing outcomes of stylet-driven versus lumenless lead implantation for LBBP are limited. In this study, we compare success rates and outcomes of LBBP with stylet-driven versus lumenless lead delivery systems.
Eighty-three consecutive patients (mean age 74.1 ± 11.2 years; 56 [68%] male) undergoing attempted LBBP at a single institution were identified. Cases were grouped by lead delivery systems used: stylet-driven (n = 53) or lumenless (n = 30). Baseline characteristics and procedural findings were recorded and compared between the cohorts. Intermediate term follow-up data on ventricular lead parameters were also compared.
Baseline characteristics were similar between groups. Successful LBBP was achieved in 77% of patients, with similar success rates between groups (76% in stylet-driven, 80% in lumenless, p = 0.79), and rates of adjudicated LBB capture and other paced QRS parameters were also similar. Compared with the lumenless group, the stylet-driven group had significantly shorter procedure times (90 ± 4 vs. 112 ± 31 min, p = 0.004) and fluoroscopy times (10 ± 5 vs. 15 ± 6 min, p = 0.003). Ventricular lead parameters at follow-up were similar, and rates of procedural complications and need for lead revision were low in both groups.
Delivery systems for stylet-driven and for lumenless leads for LBBP have comparable acute success rates. Long-term follow-up of lead performance following use of the various delivery systems is warranted. |
---|---|
ISSN: | 1045-3873 1540-8167 |
DOI: | 10.1111/jce.15789 |