High-resolution distance meter using optical intensity modulation at 28 GHz
A high resolution distance meter has been fabricated using optical waveguide modulators. The meter was evaluated in the National Research Laboratory of Metrology (Japan) semi-underground tunnel. The measurement resolution is 1 and 3 \#m\m for a distance of 5 and 100 m, respectively, in one standard...
Gespeichert in:
Veröffentlicht in: | Measurement science & technology 1998-07, Vol.9 (7), p.1049-1052 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A high resolution distance meter has been fabricated using optical waveguide modulators. The meter was evaluated in the National Research Laboratory of Metrology (Japan) semi-underground tunnel. The measurement resolution is 1 and 3 \#m\m for a distance of 5 and 100 m, respectively, in one standard deviation after correction of the cyclic error. Cyclic errors are examined for 28 GHz optical intensity modulation. (Original abstract - amended) |
---|---|
ISSN: | 0957-0233 1361-6501 |
DOI: | 10.1088/0957-0233/9/7/007 |