Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size‐specific dose estimates with position correction
This study aims to investigate the effects of the position correction of size‐specific dose estimates (SSDE) on patient dose estimation in cone beam computed tomography (CBCT). The relationship between the phantom position and absorbed dose in the right breast was studied using optically stimulated...
Gespeichert in:
Veröffentlicht in: | Journal of Applied Clinical Medical Physics 2022-12, Vol.23 (12), p.e13851-n/a |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study aims to investigate the effects of the position correction of size‐specific dose estimates (SSDE) on patient dose estimation in cone beam computed tomography (CBCT). The relationship between the phantom position and absorbed dose in the right breast was studied using optically stimulated luminescence dosimeters and a simulated human body phantom. The effect of position correction for CT dose index (CTDI) on SSDE was investigated in 51 patients who underwent right breast irradiation by comparing the SSDE with position correction and SSDE without position correction. The absorbed dose in the right breast tended to decrease by 10.2% as the phantom was placed away from the center of CBCT. The mean and standard deviation of SSDE were 2.54 ± 0.29 and 2.92 ± 0.30 mGy with and without position correction, respectively. The SSDE with position correction was 13.1% lower than that without position correction (p |
---|---|
ISSN: | 1526-9914 1526-9914 |
DOI: | 10.1002/acm2.13851 |