Stability of Hemoglobin Constant Spring Identified by Capillary Electrophoresis

Abstract Objective Hemoglobin Constant Spring (HbCS) is often missed by routine hemoglobin analysis. The aim of this research was to study HbCS stability as identified by capillary electrophoresis (CE) to determine the specimen storage time limit. Methods The EDTA blood of 29 HbCS samples were kept...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Laboratory medicine 2023-05, Vol.54 (3), p.e91-e94
Hauptverfasser: Jaitheang, Jidapa, Suksusut, Amornchai, Settapiboon, Rung, Amornsiriwat, Supaporn, Sutcharitchan, Pranee, Uaprasert, Noppacharn, Rojnuckarin, Ponlapat
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Objective Hemoglobin Constant Spring (HbCS) is often missed by routine hemoglobin analysis. The aim of this research was to study HbCS stability as identified by capillary electrophoresis (CE) to determine the specimen storage time limit. Methods The EDTA blood of 29 HbCS samples were kept at 4°C and analyzed every workday until CE could not detect HbCS or until 7 weeks after blood collection. The genotypes were confirmed by multiplex polymerase chain reaction. Results The median subject age was 27 years and 10 subjects were male. The HbCS levels were stable during the first 7 days but became undetectable in 5 cases (17.2%) after 1 week. All of them were heterozygous HbCS. Longer detection times were correlated with the higher baseline HbCS levels, with a correlation coefficient of 0.582 (P  ≤ 0.001) Conclusion Routine hemoglobin typing and quantitation should be performed within 1 week after blood collection to detect low HbCS levels, especially in heterozygous HbCS.
ISSN:0007-5027
1943-7730
DOI:10.1093/labmed/lmac130