Imageless robotic-assisted total knee arthroplasty is accurate in vivo: a retrospective study to measure the postoperative bone resection and alignment
Purpose Conventional instruments for total knee arthroplasty (TKA) have limited accuracy. The occurrence of outliers can negatively influence the clinical outcome and long-term survival of the implant. Orthopaedic robotic systems were developed to increase the accuracy of implant positioning and bon...
Gespeichert in:
Veröffentlicht in: | Archives of orthopaedic and trauma surgery 2023-06, Vol.143 (6), p.3471-3479 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Conventional instruments for total knee arthroplasty (TKA) have limited accuracy. The occurrence of outliers can negatively influence the clinical outcome and long-term survival of the implant. Orthopaedic robotic systems were developed to increase the accuracy of implant positioning and bone resections. Several systems requiring preoperative imaging have shown a higher degree of precision compared to conventional instrumentation. An imageless system needs less preoperative time and preparation and is more cost effective. Aim of this study was to find out whether this system is as precise, reproduces accurately the surgeon's planning and reduces the occurrence of outliers.
Methods
This retrospective study included the first 71 robotic-assisted TKA and 308 conventional TKA in 374 patients. Intraoperatively planned and actual bone resections were compared. Postoperative alignment, measured on full leg weight bearing radiographs, was related to the respective planning and statistically compared between the groups.
Results
Baseline characteristics (age, BMI, ASA, preoperative Knee Society Score and deformity) between both groups were comparable. According to the planned alignment, the postoperative mean difference was – 1.01° in the robotic versus 2.05° in the conventional group. The maximum deviation was − 2/+ 2.5° in the robotic and – 6.6/ + 6.8° in the conventional group. According to the plan, there were no outliers above ± 3° in the robotic versus 24% in the conventional group.
The mean difference between planned and measured bone resection was 0.21 mm with a maximum of 2 mm. The 95% confidence interval was at each position 1 mm or below.
Conclusions
The described imageless robotic system is accurate in terms of coronal alignment and bone resections. In precision, it is superior to conventional instrumentation and could therefore be used to evaluate new alignment concepts. |
---|---|
ISSN: | 1434-3916 0936-8051 1434-3916 |
DOI: | 10.1007/s00402-022-04648-2 |