Larger-diameter trunnions and bolt-reinforced taper junctions are associated with less tribocorrosion in reverse total shoulder arthroplasty
Morse taper junction tribocorrosion is recognized as an important failure mode in total hip arthroplasty. Although taper junctions are used in almost all shoulder arthroplasty systems currently available in the United States, with large variation in design, limited literature has described comparabl...
Gespeichert in:
Veröffentlicht in: | Journal of shoulder and elbow surgery 2023-01, Vol.32 (1), p.201-212 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Morse taper junction tribocorrosion is recognized as an important failure mode in total hip arthroplasty. Although taper junctions are used in almost all shoulder arthroplasty systems currently available in the United States, with large variation in design, limited literature has described comparable analyses of taper damage in these implants. In this study, taper junction damage in retrieved reverse total shoulder arthroplasty (RTSA) implants was assessed and analyzed.
Fifty-seven retrieved RTSAs with paired baseplate and glenosphere components with Morse taper junctions were identified via database query; 19 of these also included paired humeral stems and trays or spacers with taper junctions. Components were graded for standard damage modes and for fretting and corrosion with a modified Goldberg-Cusick classification system. Medical records and preoperative radiographs were reviewed. Comparative analyses were performed assessing the impact of various implant, radiographic, and patient factors on taper damage.
Standard damage modes were commonly found at the evaluated trunnion junctions, with scratching and edge deformation damage on 76% and 46% of all components, respectively. Fretting and corrosion damage was also common, observed on 86% and 72% of baseplates, respectively, and 23% and 40% of glenospheres, respectively. Baseplates showed greater moderate to severe (grade ≥ 3) fretting (43%) and corrosion (27%) damage than matched glenospheres (fretting, 9%; corrosion, 13%). Humeral stems showed moderate to severe fretting and corrosion on 28% and 30% of implants, respectively; matched humeral trays or spacers showed both less fretting (14%) and less corrosion (17%). On subgroup analysis, large-tapered implants had significantly lower summed fretting and corrosion grades than small-tapered implants (P |
---|---|