Measuring wavelengths and lattice constants with the Mössbauer wavelength standard
The newly proposed atomic‐scale length standard, the wavelength of the 57Fe Mössbauer radiation [Shvyd'ko et al. (2000). Phys. Rev. Lett.45, 495–498], is used to measure the wavelengths of the Mössbauer radiation of 151Eu, 57.556185 (27) pm, 119Sn, 51.920811 (39) pm, and 161Dy, 48.334336 (19) p...
Gespeichert in:
Veröffentlicht in: | Journal of synchrotron radiation 2002-01, Vol.9 (1), p.17-23 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The newly proposed atomic‐scale length standard, the wavelength of the 57Fe Mössbauer radiation [Shvyd'ko et al. (2000). Phys. Rev. Lett.45, 495–498], is used to measure the wavelengths of the Mössbauer radiation of 151Eu, 57.556185 (27) pm, 119Sn, 51.920811 (39) pm, and 161Dy, 48.334336 (19) pm, with a relative accuracy of ∼0.5 p.p.m. Also, the lattice constants of Al2O3 are measured in a temperature range from 286 K to 374 K. At room temperature, = 295.65 K, their values are = 4.759213 (8) Å, = 12.991586 (4) Å. |
---|---|
ISSN: | 1600-5775 0909-0495 1600-5775 |
DOI: | 10.1107/S0909049501019203 |