Evaluation of the fretting corrosion behavior of the proximal pad taper of a modular hip design

Fretting corrosion tests were conducted on Ti–6Al–4V modular `pad taper' test specimens of the Richards Modular Hip System (RMHS) in an environment containing deionized water. The primary objective was to evaluate and characterize the fretting debris production at the taper junction between the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Wear 1999-06, Vol.231 (1), p.54-64
Hauptverfasser: Chandrasekaran, V, Sauer, W.L, Taylor, A.M, Hoeppner, D.W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Fretting corrosion tests were conducted on Ti–6Al–4V modular `pad taper' test specimens of the Richards Modular Hip System (RMHS) in an environment containing deionized water. The primary objective was to evaluate and characterize the fretting debris production at the taper junction between the stem body and the proximal pad components of the RMHS by measuring metal (Ti) ion concentrations in solution during fretting tests. The fretting test parameters used in the research program were intended to simulate the hip prosthesis loading conditions. The loading parameters were: cyclic fatigue load=2.0 kN ( R=0.1), normal load=0.2 kN. These loads were selected based on standardized ISO hip stem loading conditions which in turn were based on anticipated in vivo loading conditions. Ten RMHS pad taper fretting tests were performed. Each test was performed for 5 million fretting cycles at a frequency of 6 Hz. A solution sample was removed every 1 million cycles without interrupting the fretting test. Each solution sample was analyzed for Ti ions in parts-per-million (ppm). The solution analysis test results indicated that in six cases there were no detectable ions (
ISSN:0043-1648
1873-2577
DOI:10.1016/S0043-1648(99)00110-6