HgI2 polycrystalline films for digital X-ray imagers
This paper describes recent results obtained with mercuric iodide (HgI2) polycrystalline films that we have produced. The ultimate goal of this effort is to develop a new detector technology for digital X-ray imaging based on HgI2 polycrystalline films coupled to large-area flat-panel amorphous sili...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on nuclear science 2002-02, Vol.49 (1), p.160-1164 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper describes recent results obtained with mercuric iodide (HgI2) polycrystalline films that we have produced. The ultimate goal of this effort is to develop a new detector technology for digital X-ray imaging based on HgI2 polycrystalline films coupled to large-area flat-panel amorphous silicon, thin-film transistor-addressed readout arrays. We have employed two approaches for producing the polycrystalline films: 1) thermal evaporation (sublimation) and 2) deposition of films from various solutions. The 50- to 150-micron-thick films were characterized with respect to their electrical properties and in response to ionizing radiation. The leakage current was about 40 pA/sq cm at an operating bias voltage of about 50 V. Signals from the HgI2 polycrystalline detectors, in response to ionizing radiation, compare favorably to the best published results for all high Z polycrystalline films grown elsewhere, including TlBr, PbI2, and HgI2. (Author) |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/TNS.2002.998745 |