Determining the direction of a geometrical/optical reference axis in the coordinate system of a triaxial magnetometer sensor

The reference coordinate axes of a magnetic vector field sensor are related to the instrument digital output vector by the calibration matrix and the offset vector O. In addition, this reference coordinate system must be related to (at least) two externally accessible optical or geometrical axes in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement science & technology 2002-12, Vol.13 (12), p.2094-2098
Hauptverfasser: Primdahl, F, Brauer, P, Merayo, J M G, Petersen, J R, Risbo, T
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The reference coordinate axes of a magnetic vector field sensor are related to the instrument digital output vector by the calibration matrix and the offset vector O. In addition, this reference coordinate system must be related to (at least) two externally accessible optical or geometrical axes in order to be able to determine the precise orientation of the magnetic coordinate axes in an external reference system. Two methods for determining a reference axis in the sensor coordinates are discussed: (1) a triaxial coil facility is used to measure the sensor orientation for two different positions, rotated about a fixed reference axis; (2) in the Earth's field the magnetometer sensor is rotated about a fixed axis into a number of (at least three) positions. (Author)
ISSN:0957-0233
1361-6501
DOI:10.1088/0957-0233/13/12/339