Accuracy of robotic-assisted pedicle screw placement comparing junior surgeons with expert surgeons: Can junior surgeons place pedicle screws as accurately as expert surgeons?

The purpose of this study was to verify whether a spine robotic system was useful for junior surgeons. Twenty-seven patients underwent posterior spinal fusion with open surgery using a spine robotic system (Mazor X Stealth Edition, Medtronic Inc., Dublin, Ireland) from April to August 2021. Pedicle...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association 2023-09, Vol.28 (5), p.961-965
Hauptverfasser: Torii, Yoshiaki, Ueno, Jun, Iinuma, Masahiro, Yoshida, Atsuhiro, Niki, Hisateru, Akazawa, Tsutomu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to verify whether a spine robotic system was useful for junior surgeons. Twenty-seven patients underwent posterior spinal fusion with open surgery using a spine robotic system (Mazor X Stealth Edition, Medtronic Inc., Dublin, Ireland) from April to August 2021. Pedicle screw insertions were performed by five surgeons. The surgeon and insertion time were recorded for each pedicle screw. Two surgeons who are board-certified spine surgeons by the Japanese Society for Spine Surgery and Related Research (JSSR) were defined as the expert surgeon group. Three surgeons who were training to acquire qualifications as JSSR board certified spine surgeons were defined as the junior surgeon group. In postoperative CT images, the deviation of 255 pedicle screws was evaluated using the Gertzbein-Robbins (GR) grades. In the expert surgeon group, the GR grades were Grade A for 79 screws (90.8%), Grade B for 6 (6.9%), Grade C for 2 (2.3%), and 0 (0%) for Grades D and E. I In the junior surgeon group, the GR grades were Grade A for 162 screws (96.4%), Grade B for 6 (3.6%), and 0 (0%) for Grades C, D, and E. There was no significant difference in the deviation rate between surgeon groups (p = 0.08). The mean insertion times were 174.5 ± 83.0 s in the expert surgeon group and 191.0 ± 111.0 s in the junior surgeon group. There was no significant difference in the insertion time between surgeon groups (p = 0.22). There were no significant differences in the deviation rate and the insertion time of robotic-assisted pedicle screw placement between expert surgeons and junior surgeons who were training to acquire qualifications as JSSR board certified spine surgeons. Robotic-assisted pedicle screw placement can be effectively employed by junior surgeons.
ISSN:0949-2658
1436-2023
DOI:10.1016/j.jos.2022.06.012