Deformation of Transcatheter Aortic Valve Prostheses: Implications for Hypoattenuating Leaflet Thickening and Clinical Outcomes

Although transcatheter aortic valve replacement (TAVR) therapy continues to grow, there have been concerns about the occurrence of hypoattenuating leaflet thickening (HALT), which may affect prosthesis function or durability. This study aimed to examine prosthesis frame factors and correlate their e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Circulation (New York, N.Y.) N.Y.), 2022-08, Vol.146 (6), p.480-493
Hauptverfasser: Fukui, Miho, Bapat, Vinayak N., Garcia, Santiago, Dworak, Marshall W., Hashimoto, Go, Sato, Hirotomo, Gössl, Mario, Enriquez-Sarano, Maurice, Lesser, John R., Cavalcante, João L., Sorajja, Paul
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Although transcatheter aortic valve replacement (TAVR) therapy continues to grow, there have been concerns about the occurrence of hypoattenuating leaflet thickening (HALT), which may affect prosthesis function or durability. This study aimed to examine prosthesis frame factors and correlate their extent to the frequency of HALT and clinical outcomes. We prospectively examined 565 patients with cardiac computed tomography screening for HALT at 30 days after balloon-expandable SAPIEN3 and self-expanding EVOLUT TAVR. Deformation of the TAVR prostheses, asymmetric prosthesis leaflet expansion, prosthesis sinus volumes, and commissural alignment were analyzed on the postprocedural computed tomography. For descriptive purposes, an index of prosthesis deformation was calculated, with values >1.00 representing relative midsegment underexpansion. A time-to-event model was performed to evaluate the association of HALT with the clinical outcome. Overall, HALT was present in 21% of SAPIEN3 patients and in 16% of EVOLUT patients at 30 days after TAVR. The occurrence of HALT was directly associated with greater prosthesis frame deformation (
ISSN:0009-7322
1524-4539
DOI:10.1161/CIRCULATIONAHA.121.058339