The impact of Morse taper implant design on microleakage at implant-healing abutment interface

The microbial leakage at the implant–abutment interface (IAI) is one cause of peri-implant infection that puts long-term implant stability at risk. The present study compared two types of Morse taper implants in terms of sealing performance at the implanthealing abutment interface. Three implant sys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dental Materials Journal 2022/09/25, Vol.41(5), pp.767-773
Hauptverfasser: KIM, Soyeon, LEE, Joo Won, KIM, Jae-Heon, TRUONG, Van Mai, PARK, Young-Seok
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The microbial leakage at the implant–abutment interface (IAI) is one cause of peri-implant infection that puts long-term implant stability at risk. The present study compared two types of Morse taper implants in terms of sealing performance at the implanthealing abutment interface. Three implant systems, one of which exhibited a partial face-contact design (TSO) and the other two a line-contact design (TSM and BLT), were analyzed in vitro using two quantitative implant leakage measuring techniques: air-injection pressure measurement test and microbial examination. An in-house-developed device was used to determine implant leakage by recording the initial drop in pressure while injecting air through the implant. The microbial examination measured the absorbances of culture mediums in which the inoculated implants were contained. Significant difference was found between the partial face- and line-contact groups (p
ISSN:0287-4547
1881-1361
DOI:10.4012/dmj.2022-022