Negligible Specimen Hemolysis Is Observed With Sarstedt S-Monovette Coagulation Tubes in Aspiration Mode
Abstract Objectives We aimed to compare the levels of hemolysis in the blood collected using the vacuum and aspiration modes via Sarstedt S-Monovette coagulation tubes. Methods Forty volunteers were included in the study. Blood samples were collected using two different modes in the S-Monovette citr...
Gespeichert in:
Veröffentlicht in: | American journal of clinical pathology 2022-10, Vol.158 (4), p.494-498 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Objectives
We aimed to compare the levels of hemolysis in the blood collected using the vacuum and aspiration modes via Sarstedt S-Monovette coagulation tubes.
Methods
Forty volunteers were included in the study. Blood samples were collected using two different modes in the S-Monovette citrate tube (Sarstedt AG). Prothrombin time, active partial thromboplastin time, fibrinogen, and D-dimer analyses were performed using the STA-Compact-Max 3 analyzer (Stago). The hemolysis levels of the samples were measured by both Stago’s semiquantitative hemolysis index (H-index) module and the quantitative H-index measurement of the Roche cobas 6000 (Roche Diagnostics) analyzer.
Results
Roche’s quantitative H-index values were statistically significantly lower in the aspiration mode. No clinically significant difference was observed between coagulation test results.
Conclusions
Using the S-Monovette citrate tubes can reduce spurious hemolysis and improve patient safety. |
---|---|
ISSN: | 0002-9173 1943-7722 |