Enzyme sensor for simultaneous determination of cholesterol, triglyceride and HDL cholesterol in serum
A three-electrode lipid biosensor that simultaneously measures the total cholesterol (CHO), triglyceride (TG), and HDL cholesterol (HDL-C) in standard serum has been developed. The lipid biosensor is designed for clinical use where production cost, low sample requirement, portability, stability, and...
Gespeichert in:
Veröffentlicht in: | Analytical sciences 2022-09, Vol.38 (9), p.1189-1197 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A three-electrode lipid biosensor that simultaneously measures the total cholesterol (CHO), triglyceride (TG), and HDL cholesterol (HDL-C) in standard serum has been developed. The lipid biosensor is designed for clinical use where production cost, low sample requirement, portability, stability, and speed are high priorities. The device design filters out blood cells and lipoproteins from the serum, where the target molecules are catalyzed by enzymes encapsulated in carboxymethyl cellulose (CMC) to produce electron mediators, potassium ferrocyanide. These electron mediators were subsequently detected by amperometric determination. The sensor exhibit high selectivity towards the targets and can measure the target lipids in 4 min with 10 µL of serum. Lastly, the device can be stored up to 18 months with a minimal decrease in catalytic efficiency.
Graphical abstract |
---|---|
ISSN: | 0910-6340 1348-2246 |
DOI: | 10.1007/s44211-022-00148-w |