Talo-navicular and calcaneo-cuboid fusion with PEEK H-pode™ vs. Titanium Maxlock™ locking plates: Comparison of functional and radiographic outcomes
Locking plates are increasingly used to achieve hindfoot fusion. The objective of this study was to compare hindfoot fusion outcomes with the PEEK H-pode™ (Biotech™) locking plate and the titanium Maxlock™ (Tornier-Wright™) locking plate. A polyetheretherketone (PEEK) H-pode™ locking plate provides...
Gespeichert in:
Veröffentlicht in: | Orthopaedics & traumatology, surgery & research surgery & research, 2023-02, Vol.109 (1), p.103343-103343, Article 103343 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Locking plates are increasingly used to achieve hindfoot fusion. The objective of this study was to compare hindfoot fusion outcomes with the PEEK H-pode™ (Biotech™) locking plate and the titanium Maxlock™ (Tornier-Wright™) locking plate.
A polyetheretherketone (PEEK) H-pode™ locking plate provides similar fusion rates to a titanium Maxlock™ locking plate for talo-navicular and calcaneo-cuboid arthrodesis.
We conducted a retrospective comparative study in 39 patients (including 21 [54%] with pes planovalgus and 11 [28%] with neurological equinovarus deformities) who underwent talo-navicular and/or calcaneo-cuboid fusion, usually combined with subtalar fusion. The first 17 patients (January 2014–February 2016) were managed with Maxlock™ locking plates and the next 22 patients (March 2016–August 2018) with H-pode™ locking plates. These two cohorts of consecutive patients were comparable regarding age, sex distribution, body mass index, and comorbidities. At last follow-up more than 1year after surgery, we compared functional scores, pain intensity, and fusion assessed by radiographs and computed tomography (CT) (threshold set at 33%).
Mean follow-up was 42months (range: 34–63months) in the Maxlock™ group and 25months (range: 12–36months) in the H-pode™ group. At last follow-up, the two groups were not significantly different for the mean values of the American Orthopedic Foot & Ankle Society score, European Foot & Ankle Society score, and visual analogue scale pain score. Neither were the two groups significantly different for talo-navicular and calcaneo-cuboid fusion by CT. In the Maxlock™ group, we found non-significant trends towards a higher proportion of patients with talo-navicular nonunion (18% vs. 5% in the H-pode™ group) and weaker talo-navicular fusion in patients with pes planovalgus (60.8% vs. 82.0%, respectively). Radiographic results overestimated the fusion rates.
Talo-navicular and calcaneo-cuboid fusion was not significantly different with H-pode™ and Maxlock™ locking plates. CT was more accurate than standard radiography to assess fusion. These results underline the usefulness of PEEK locking plates for talo-navicular and calcaneo-cuboid fusion; moreover, the radiolucency of PEEK facilitates the interpretation of radiographs.
IV. |
---|---|
ISSN: | 1877-0568 1877-0568 |
DOI: | 10.1016/j.otsr.2022.103343 |