Point-of-Care: Roadmap for Analytical Characterization and Validation of a High-Sensitivity Cardiac Troponin I Assay in Plasma and Whole Blood Matrices
High-sensitivity cardiac troponin (hs-cTn) assays enable more precise use of traditional diagnostic strategies and earlier rule-out/rule-in at 0/1 h or 0/2 h after presentation of acute myocardial infarction (AMI). Availability of hs-cTn measurements at point-of-care (POC) can improve timely managem...
Gespeichert in:
Veröffentlicht in: | The journal of applied laboratory medicine 2022-06, Vol.7 (4), p.971-988 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High-sensitivity cardiac troponin (hs-cTn) assays enable more precise use of traditional diagnostic strategies and earlier rule-out/rule-in at 0/1 h or 0/2 h after presentation of acute myocardial infarction (AMI). Availability of hs-cTn measurements at point-of-care (POC) can improve timely management of AMI patients. A roadmap for regulatory and analytical validation is exemplified with studies with the Atellica VTLi hs-cTnI at POC.
High-sensitivity performance was assessed with AACC/IFCC expert recommendations. Clinical Laboratory Standards Institute protocols were used for characterizing limit of blank, limit of detection (LoD), limit of quantitation (LoQ), 10% CV, precision, linearity, and analytic specificity with several reagent lots. Bland-Altman, Passing-Bablok, and hematocrit bias plots compared hs-cTnI measurement in lithium-heparin plasma (PL) and whole blood (WB) matrices.
LoB was 0.55 ng/L; LoD and LoQ were 1.24 ng/L and 2.1 ng/Lm for PL and 1.60 ng/L and 3.7 ng/L for WB, respectively. The male 99th percentile is 27 ng/L, and female 99th percentile upper reference limit is 18 ng/L; 10% CVs were 6.7 ng/L for PL and 8.9 ng/L for WB. Also ≥50% of hs-cTnI values for healthy cohorts exceeded the LoD, confirming high-sensitivity performance. Linearity spanned from LoQ to 1250 ng/L. Specificity was >90% for 40 potential interferences; no hook effect was detected. WB and PL correlation was WB = 1.02*plasma + 0.3 ng/L (r = 0.996, n = 152). No hs-cTnI association with hematocrit was detected (R2 = 0.003).
This analytical roadmap showed high-sensitivity performance, good analytic characteristics, and excellent PL and WB agreement for the Atellica VTLi hs-cTnI POC system. Essential clinical performance studies in patients by intended POC users may now commence. |
---|---|