Aneurysmal recurrence after successful flow-diversion embolization
Aneurysmal recurrence after successful flow-diversion embolization is exceptionally rare. The rarity of recurrence has called into question the yield of interval surveillance imaging. Here we report the case of a recurrent intracranial aneurysm despite complete angiographic resolution after flow-div...
Gespeichert in:
Veröffentlicht in: | Interventional neuroradiology 2024-04, Vol.30 (2), p.297-301 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aneurysmal recurrence after successful flow-diversion embolization is exceptionally rare. The rarity of recurrence has called into question the yield of interval surveillance imaging. Here we report the case of a recurrent intracranial aneurysm despite complete angiographic resolution after flow-diversion therapy with a Pipeline embolization device (PED). Given the absence of poor wall apposition, endoleak, and device migration, how this aneurysm recurred remains unclear, particularly given the recurrence was at a timepoint at which complete reendothelialization of the device would be expected. The patient ultimately underwent interval treatment with a second device placed across the neck of the aneurysm. Although rare, reports of aneurysmal recurrences support the use of interval non-invasive imaging surveillance to ensure successful embolization in this patient population. |
---|---|