Influence of field of view size and reconstruction methods on single-energy metal artifact reduction: a phantom study

The purpose of this study was to evaluate the effect of single-energy metal artifact reduction (SEMAR) for metal artifacts using CT images reconstructed with adaptive iterative dose reduction three dimensional (AIDR3D) and advanced intelligent clear-IQ engine (AiCE) in calibration-field of view of v...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Australasian physical & engineering sciences in medicine 2022-06, Vol.45 (2), p.637-642
Hauptverfasser: Tsuboi, Kunihito, Osaki, Naoko, Ohtani, Yuki, Tanikawa, Ken, Kaneko, Masanori
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to evaluate the effect of single-energy metal artifact reduction (SEMAR) for metal artifacts using CT images reconstructed with adaptive iterative dose reduction three dimensional (AIDR3D) and advanced intelligent clear-IQ engine (AiCE) in calibration-field of view of various sizes. A prosthetic hip joint was arranged at the center of the phantom. The phantom images were scanned by changing calibration-field of view of 320 mm and 500 mm, and were reconstructed using filtered back-projection (FBP), AIDR3D, and AiCE with and without SEMAR, respectively. The metal artifact reduction with SEMAR was evaluated by calculated the relative artifact index value and visual scores in degree of artifact by seven radiology technologists. Relative artifact index of FBP, AIDR3D, and AiCE with 320 mm/500 mm calibration-field of views were 10.2/10.0, 16.3/16.4, and 17.8/17.9 without SEMAR, 3.3/3.1, 2.6/2.5, and 2.3/2.0 with SEMAR, respectively. Visual scores were not significantly different between 320 and 500 mm calibration-field of views in all reconstruction methods. The effect of metal artifact reduction was not affected by calibration-field of view sizes in the SEMAR combined with AIDR3D or AiCE.
ISSN:2662-4729
0158-9938
2662-4737
1879-5447
DOI:10.1007/s13246-022-01130-6