Biocompatibility and antibiofilm activity of graphene-oxide functionalized titanium discs and collagen membranes

The aims of the study were: 1) to evaluate the effect on biofilm formation of barrier membranes and titanium surfaces coated with graphene-oxide (GO); 2) to analyze the connection between the superficial topography of the tested materials and the amount of bacterial accumulation on them and 3) to an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dental materials 2022-07, Vol.38 (7), p.1117-1127
Hauptverfasser: Radunovic, Milena, Pavic, Aleksandar, Ivanovic, Vera, Milivojevic, Marija, Radovic, Igor, Di Carlo, Roberta, Pilato, Serena, Fontana, Antonella, Piattelli, Adriano, Petrovic, Sanja
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aims of the study were: 1) to evaluate the effect on biofilm formation of barrier membranes and titanium surfaces coated with graphene-oxide (GO); 2) to analyze the connection between the superficial topography of the tested materials and the amount of bacterial accumulation on them and 3) to analyze the biocompatibility of GO functionalized discs using the zebrafish model. Single species bacterial biofilms (Streptococcus oralis, Veilonella parvula, Fusobacterium nucleatum, Porphyomonas gingivalis) were grown on GO-free membranes, membranes coated with 2 and 10 μg/ml of GO, GO-free and GO-coated titanium discs. The biofilms were analyzed by determining the CFU count and by Scanning Electron Microscopy (SEM) and the materials’ topography by Atomic Force Microscopy (AFM). Zebrafish model was used to determine the materials’ toxicity and inflammatory effects. AFM showed similar roughness of control and GO-coated materials. CFU counts on GO-coated discs were significantly lower than on control discs for all species. CFU counts of S. oralis, V. parvula and P. gingivalis were lower on biofilms grown on both types of GO-coated membranes than on GO-free membrane. SEM analysis showed different formation of single species biofilm of S. oralis on control and GO-coated materials. GO-functionalized titanium discs do not induce toxic or inflammatory effects. Titanium implant surfaces functionalized with GO have shown to be biocompatible and less susceptible to biofilm formation. These results encourage further in vivo investigation of the tested materials on infection prevention, specifically in prevention and reduction of peri-implant mucositis and periimplantitis incidence. [Display omitted] •GO coated materials show antibiofilm effect.•The antibiofilm effect of GO is stronger on Gram + than on Gram – bacteria.•Surface roughness of GO-coated materials does not influence biofilm formation.•Disks functionalized with GO does not show toxicity.
ISSN:0109-5641
1879-0097
DOI:10.1016/j.dental.2022.04.024