Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator

There is currently much interest in the characterisation of wear debris from different types of artificial hip joints. There have been numerous studies on the wear of UHMWPE in hip joint simulators, but relatively few studies on the wear of alternative materials such as metal-on-metal (MOM) and cera...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Wear 2001-10, Vol.250 (1-12), p.120-128
Hauptverfasser: Tipper, J.L, Firkins, P.J, Besong, A.A, Barbour, P.S.M, Nevelos, J, Stone, M.H, Ingham, E, Fisher, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:There is currently much interest in the characterisation of wear debris from different types of artificial hip joints. There have been numerous studies on the wear of UHMWPE in hip joint simulators, but relatively few studies on the wear of alternative materials such as metal-on-metal (MOM) and ceramic-on-ceramic (COC). The aim of this study was to compare the wear volumes and wear debris generated from zirconia ceramic-on-UHMWPE, MOM and COC hip joints under identical conditions in the same hip joint simulator. All prostheses showed an initial higher ‘bedding in’ wear rate, which was followed by a lower steady state wear rate. The zirconia ceramic-on UHMWPE prostheses showed the highest wear rates (31±4.0mm3/million cycles), followed by the MOM (1.23±0.5mm/million cycles), with the COC prostheses showing significantly (P
ISSN:0043-1648
1873-2577
DOI:10.1016/S0043-1648(01)00653-6