Accuracy of a Factory-Calibrated Continuous Glucose Monitor in Individuals With Diabetes on Hemodialysis

Continuous glucose monitoring (CGM) improves diabetes management, but its reliability in individuals on hemodialysis is poorly understood and potentially affected by interstitial and intravascular volume variations. We assessed the accuracy of a factory-calibrated CGM by using venous blood glucose m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Diabetes care 2022-07, Vol.45 (7), p.1666-1669
Hauptverfasser: Villard, Orianne, Breton, Marc D, Rao, Swati, Voelmle, Mary K, Fuller, Morgan R, Myers, Helen E, McFadden, Ryan K, Luke, Zander S, Wakeman, Christian A, Clancy-Oliveri, Mary, Basu, Ananda, Stumpf, Meaghan M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Continuous glucose monitoring (CGM) improves diabetes management, but its reliability in individuals on hemodialysis is poorly understood and potentially affected by interstitial and intravascular volume variations. We assessed the accuracy of a factory-calibrated CGM by using venous blood glucose measurements (vBGM) during hemodialysis sessions and self-monitoring blood glucose (SMBG) at home. Twenty participants completed the protocol. The mean absolute relative difference of the CGM was 13.8% and 14.4%, when calculated on SMBG (n = 684) and on vBGM (n = 624), and 98.7% and 100% of values in the Parkes error grid A/B zones, respectively. Throughout 181 days of CGM monitoring, the median time in range (70-180 mg/dL) was 38.5% (interquartile range 29.3-57.9), with 28.7% (7.8-40.6) of the time >250 mg/dL. The overall performance of a factory-calibrated CGM appears reasonably accurate and clinically relevant for use in practice by individuals on hemodialysis and health professionals to improve diabetes management.
ISSN:0149-5992
1935-5548
DOI:10.2337/dc22-0073