A phantom study using dual-energy spectral computed tomography imaging: Comparison of image quality between two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms for evaluating ground-glass nodules of the lung
Aims: To assess the image quality of the two adaptive statistical iterative reconstruction (ASiR and ASiR-V) algorithms for evaluating ground-glass nodules (GGNs) of the lung. Subjects and Methods: The chest phantom with ground-glass simulated pulmonary nodules was scanned by dual-energy spectral co...
Gespeichert in:
Veröffentlicht in: | Journal of cancer research and therapeutics 2021-12, Vol.17 (7), p.1742-1747 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aims: To assess the image quality of the two adaptive statistical iterative reconstruction (ASiR and ASiR-V) algorithms for evaluating ground-glass nodules (GGNs) of the lung.
Subjects and Methods: The chest phantom with ground-glass simulated pulmonary nodules was scanned by dual-energy spectral computed tomography (CT) using gemstone spectral imaging mode. The image was reconstructed with ASiR and ASiR-V from 0 (FBP) to 90% blending percentages, respectively. The average noise and signal-to-noise ratio (SNR) of reconstruction images were calculated. The data were statistically analyzed.
Results: Compared with FBP, as the percentage of ASiR and ASiR-V increased from 10 to 90%, image noise reduced by 3.7%-45.2% and 14.1%-73.8%, respectively, and the SNR increased accordingly. There was significantly higher SNR value of ASiR-V images as the percentages of IR increased to 50% or greater, compared to those of ASiR. Subjective image quality scores of ASiR and ASiR-V improved significantly as percentage increased from 10 to 80% for ASiR and ASiR-V (peaked at 60% for both of them).
Conclusions: Both ASiR and ASiR-V can reduce the image noise and improve the objective image quality for presenting pulmonary nodules on dual-energy spectral CT imaging compared with FBP. ASiR-V provided further image quality improvement than ASiR, and both ASiR and ASiR-V 60% had the highest image quality score. |
---|---|
ISSN: | 0973-1482 1998-4138 |
DOI: | 10.4103/jcrt.JCRT_1780_20 |