Reliability of Bioreactance and Pulse-Power Analysis in Measuring Cardiac Index in Patients Undergoing Cardiac Surgery With Cardiopulmonary Bypass

Less-invasive and continuous cardiac output monitors recently have been developed to monitor patient hemodynamics. The aim of this study was to compare the accuracy, precision, and trending ability of noninvasive bioreactance-based Starling SV and miniinvasive pulse-power device LiDCOrapid to bolus...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of cardiothoracic and vascular anesthesia 2022-08, Vol.36 (8), p.2446-2453
Hauptverfasser: Ylikauma, Laura Anneli, Lanning, Katriina Marjatta, Erkinaro, Tiina Maria, Ohtonen, Pasi Petteri, Vakkala, Merja Annika, Liisanantti, Janne Henrik, Juvonen, Tatu Sakari, Kaakinen, Timo Ilari
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Less-invasive and continuous cardiac output monitors recently have been developed to monitor patient hemodynamics. The aim of this study was to compare the accuracy, precision, and trending ability of noninvasive bioreactance-based Starling SV and miniinvasive pulse-power device LiDCOrapid to bolus thermodilution technique with a pulmonary artery catheter (TDCO) when measuring cardiac index in the setting of cardiac surgery with cardiopulmonary bypass (CPB). A prospective method-comparison study. Oulu University Hospital, Finland. Twenty patients undergoing cardiac surgery with CPB. Cardiac index measurements were obtained simultaneously with TDCO intraoperatively and postoperatively, resulting in 498 measurements with Starling SV and 444 with LiDCOrapid. The authors used the Bland-Altman method to investigate the agreement between the devices and four-quadrant plots with error grids to assess the trending ability. The agreement between TDCO and Starling SV was qualified with a bias of 0.43 L/min/m2 (95% confidence interval [CI], 0.37-0.50), wide limits of agreement (LOA, –1.07 to 1.94 L/min/m2), and a percentage error (PE) of 66.3%. The agreement between TDCO and LiDCOrapid was qualified, with a bias of 0.22 L/min/m2 (95% CI 0.16-0.27), wide LOA (–0.93 to 1.43), and a PE of 53.2%. With both devices, trending ability was insufficient. The reliability of bioreactance-based Starling SV and pulse-power analyzer LiDCOrapid was not interchangeable with TDCO, thus limiting their usefulness in cardiac surgery with CPB.
ISSN:1053-0770
1532-8422
DOI:10.1053/j.jvca.2021.11.039